Iraki, J, Paulsen, G, Garthe, I, Slater, G and Areta, JL (2021) Reliability of resting metabolic rate between and within day measurements using the Vyntus CPX system and comparison against predictive formulas. Nutrition and Health. ISSN 2047-945X
|
Text
Reliability of resting metabolic rate between and within day measurements.pdf - Published Version Available under License Creative Commons Attribution Non-commercial. Download (969kB) | Preview |
Abstract
Background: To detect longitudinal changes of resting metabolic rate (RMR) resulting from the effects of energetic stress, reliable RMR measurements are crucial. The Vyntus CPX is a new automated indirect calorimetry system for which RMR reliability has not been determined. Additionally, its agreement with common predictive RMR formulas is unknown. Aim: To determine the within and between-day reliability of RMR measurements using the Vyntus CPX system and its agreement with predictive RMR formulas. Methods: Young (31 ± 7 years) healthy participants (n = 26, 12 females, 14 males) completed three measurements of RMR, two consecutive measures on the same day, one the day before/after, all under standardised conditions. Reliability was assessed with pairwise comparisons of between-day at the same time (BDST), within day consecutive measurements (WDCM) and between-day different time (BDDT), for parameters of reliability (mean change (MC), intraclass correlation (ICC) and typical error of measurement (TEM)). Measured RMR values (kcal/day) were compared against predictive values of 4 common formulas. Results: Parameters of reliability (mean, (95% confidence interval)) were: -BDST: MC, 0.2(-2.3—2.7)% (p = 0.67); ICC, 0.92(0.84—0.97); TEM, 4.5(3.5—6.2)%. -WDCM: MC, −2.5(-6.2—1.3)% (p = 0.21); ICC, 0.88(0.74—0.88); TEM, 7.0(5.4—9.8)%. -BDDT: MC, −1.5(-4.8—1.9)% (p = 0.57); ICC, 0.90(0.76—0.95); TEM, 6.1(4.8—8.5)%. RMRratios (measured/predicted) were: 1.04 ± 0.14 (Nelson, p = 0.13), 1.03 ± 0.10 (Mifflin, p = 0.21), 0.98 ± 0.09 (Harris-benedict, p = 0.30), 0.95 ± 0.11 (Cunningham1980, p = 0.01), 1.00 ± 0.12 (Cunningham1991, p = 0.90) and 0.96 ± 0.13 (DXA, p = 0.03). Conclusions: The Vyntus CPX is reliable and measured RMR values agreed with four predictive formulas but are lower than Cunningham1980 and DXA RMR estimates for this population.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | RMRratio; indirect calorimetry; reliability; resting metabolic rate; Nutrition & Dietetics |
Subjects: | R Medicine > RC Internal medicine > RC1200 Sports Medicine |
Divisions: | Sport & Exercise Sciences |
Publisher: | SAGE Publications |
SWORD Depositor: | A Symplectic |
Date Deposited: | 31 Jan 2023 12:30 |
Last Modified: | 01 Feb 2023 09:56 |
DOI or ID number: | 10.1177/02601060211057324 |
URI: | https://researchonline.ljmu.ac.uk/id/eprint/18775 |
View Item |